Red & Blue Light Phototherapy Device

  • Purchase Qty: 1
  • Unit: Vial(s)
  • Fob Unit Price: US $7200
  • Payment Terms:L/C , T/T
  • Favorite
Yunnan Original Color Biomedical Technology Co., Ltd.

Business Type:Trader

Country/Region:China

Ddu Verified

HOT Rank

8/10

Product Information

  • Certification:CE
  • Place of Origin:China
  • Warranty:10 years

Description

  1. Power Supply: AC 220V / 50Hz; Input Power: 370VA
  2. LCD Display: 10.1-inch touch screen for real-time display of treatment parameters
  3. Channels: Dual-channel, equipped with two large lamp heads
  4. Wavelength: Red light: 630nm±10nm; Blue light: 465nm±10nm

★5. Irradiance on the surface of irradiation window a) Red light: Level 1: 120mW/cm², Level 2: 450mW/cm², Level 3: 800mW/cm²; tolerance ±20% b) Blue light: Level 1: 120mW/cm², Level 2: 500mW/cm², Level 3: 1200mW/cm²; tolerance ±20% c) Mixed light: Level 1: 240mW/cm², Level 2: 950mW/cm², Level 3: 1900mW/cm²; tolerance ±20%

★6. Irradiance on effective irradiation area a) Red light: Level 1: 40mW/cm², Level 2: 80mW/cm², Level 3: 120mW/cm²; tolerance ±20% b) Blue light: Level 1: 50mW/cm², Level 2: 100mW/cm², Level 3: 150mW/cm²; tolerance ±20% c) Mixed light: Level 1: 90mW/cm², Level 2: 180mW/cm², Level 3: 270mW/cm²; tolerance ±20%

  1. Uniformity of effective irradiance shall be greater than 0.4
  2. Instability of effective irradiance shall not exceed ±10%
  3. Radiation spectrum: The ratio of irradiance within 600nm~760nm to irradiance within 200nm~1400nm shall be no less than 0.8
  4. Timer a) Adjustable from 5min~60min with 5min step increment, tolerance ±2%. The device will stop working immediately and trigger an audible alert once treatment completes. b) Manual light output stop function available.

★11. Temperature monitoring & over-temperature protection a) The device monitors temperature of effective irradiation area. Monitoring range: 10℃~50℃, tolerance ±3℃ b) Over-temperature protection: When the temperature of effective irradiation area exceeds 42℃, the device shall cut off light output immediately and cannot resume automatically.

  1. Cooling system a) Cooling method: Fan cooling b) The fan starts automatically when the lamp head is working to cool the light source.

Provides anti-inflammatory, bactericidal and pain-relieving effects to accelerate wound healing.


  • Dual-light synergistic therapy: red & blue light available separately or combined
  • Blue light kills Propionibacterium acnes for acne; red light penetrates dermis, boosts collagen, reduces inflammation & speeds wound healing
  • High optical power density with strong photobiochemical effects for consistent treatment results
  • Multi-indication application: acne, herpes zoster, skin ulcers, dermatitis and eczema
  • High-power semiconductor light source with specific wavelengths to act on subcutaneous tissue
  • Adjustable irradiation head & multi-axis arm for precise positioning, comfortable treatment